Module OEM Implementation Manual
Welch Allyn OEM Technologies Confidential Page 77
Appendix E
Specifications
LC101 Main Board
PHYSICAL SPECIFICATIONS
Size 3.15 in. x 3.90 in. (80.3 mm x 99.3 mm)
Clearance 0.85 in. (22 mm) above board, 0.12 in (3 mm) below board
Orientation No limitations
Weight Less than 1 pound (454 grams) with full pneumatics.
SYSTEM PERFORMANCE SPECIFICATIONS
Sampling Method Built-in sidestream sensor
Patient Application Neonatal to adult
CO
2
Concentration Display Range 0 - 99 mmHg
Respiration Rate Range 1 - 99 breaths per minute
Typical CO
2
Accuracy
a
Start-Up Time Less than 10 seconds to acquire CO
2
waveform data
Less than 3 minutes to full operating specification.
Calibration Zero Calibration and Two-Point User Calibration
C-Cap Rise Time Less than 200 msec (10% - 90%, bench only)
Flow Rate Range User-selectable, variable from 90 to 200 ml/min (defaults to 175
ml/min)
Flow Rate Accuracy -20/+15% of set value.