Module OEM Implementation Manual
Page 44 Confidential Welch Allyn OEM Technologies
Status Message Code Table
Code (hex) Suggested Message
00 Status OK
01 Invalid Command
02 Invalid Data
03 Unprotected or Unprotected Operation
06 Acknowledge Mode Command
11 CO
2
Sensor Start Up In Progress
15 Vacuum Offset too Large
16 No Watertrap
17 Watertrap Or Cannula Occlusion
18 Exhaust Occlusion Or Pneumatic Leak
21 Calibration Already in Progress
22 Calibration Not in Progress
23 Low Run Time
24 Calibration Ready For Next Step
25 Calibration in Progress
26 Calibration OK
27 Calculation Error
28 Unable to Perform Calibration, Parameters missing
29 Unable to Perform Calibration, Data Error
2A Bad Calibration CRC
40 Watchdog Error
44 System EEPROM CCITT/CRC Error
46 System FLASH ROM CCITT/CRC Error
47 System Communication Error (LC101 internal)
4B System External RAM Error
4C System RAM Error
4D System FLASH ROM Error/Checksum
4E Stack Overflow Error
4F System SW Error/Main Program
51 Manufacturer ID Mismatch
57 Sensor Not Found
60 CO
2
Sensor EEPROM Error/Revision #
65 CO
2
Sensor EEPROM Error/Read/Write
66 CO
2
Sensor EEPROM Error/CCITT/CRC Code
70 CO
2
Sensor Temperature Too High
71 CO
2
Sensor Temperature Too Low
80 Pump Failure
81 Unexpected Reverse Flow
82 Unexpected Forward Flow
84 Barometric Pressure Too High
85 Barometric Pressure Too Low