Vista de pagina 36
LC101 CO
2
Module OEM Implementation Manual
Section 4 - Host/Module Communications
Welch Allyn OEM Technologies Confidential Page 37
<C27>
Request single breath packet
<zxxyyww> xx = ETCO
2
mmHg
yy = RR bpm
ww = InsCO
2
mmHg
Example:
<z201500>
ETCO
2
= 32
RR= 21
InsCO2= 0
<C2E>
Request current pump flow rate
<Bxxxx> xxxx = pump flow rate ml/min
Example:
<B0096>
ow rate = 150 ml/min
<C31>
Request sensor EEPROM revi-
sion #
<Rxx> xx = revision #
Example:
<R02>
revision # = 2
<C32>
Request sensor customer code
<Ixx> xx = customer code
Example:
<I00>
customer code = 00
<C33>
Request sensor serial #
<Nxxxx> xxxx = serial #
Example:
<N34C2>
serial # = 13506
<C3A>
Request last calibration date
<Dmmddyyyy> mm = month
dd = day
yyyy = year
Example:
<D06041998>
last cal date= 6/4/1998