Module OEM Implementation Manual
Page 38 Confidential Welch Allyn OEM Technologies
Configuration Command Responses
Configuration responses are data packets sent by the Module to the host at programmed
intervals. A single host measurement or auto run mode command initiates unsolicited
transmission of CO
2
waveform and breath data packets. The programmed interval between
packets is either defined by the host via a configuration command or the host may use the default
setting.
<Wxxyy>, CO
2
Waveform Data
Includes 4 bytes of CO
2
waveform data that are sent at the programmed interval during
measurement and auto run modes.
<Wxxyy> where xxyy:
xxyy CO
2
waveform data mmHg × 256 4*(ASCII)
range: 0 – 99.996 mmHg (0000h – 63FFh)
Example:
CO
2
= 37.50 mmHg
Module Response Module Response Format
<Wxxyy>
Unsolicited CO
2
waveform packet at programmed interval
see below
<Zxxyyzz>
Unsolicited breath data packet at programmed interval
see below
{STX} “W” xx yy CCITT/CRC {ETX} waveform byte string
0x25 0x80 0x79 {STX}W258079{ETX}
02h 57h 32h35h 38h30h 37h39h 03h