LC101 CO
2
Module OEM Implementation Manual
Section 4 - Host/Module Communications
Welch Allyn OEM Technologies Confidential Page 39
<Zxxyyzz>, Breath Data
Includes 6 bytes of breath data that are sent at the programmed interval during measurement
and auto run modes.
<Zxxyyzz> where xxyyzz:
xx ETCO
2
mmHg 2*(ASCII)
yy RR bpm 2*(ASCII)
zz InsCO
2
mmHg 2*(ASCII)
Range:
ETCO
2
0 – 99 mmHg (00h – 63h)
RR 0 – 250 bpm (00h – FAh)
InsCO
2
0 – 99 mmHg (00h – 63h)
Example:
ETCO
2
= 39 mmHg / RR = 12 bpm / InsCO
2
= 0 mmHg
Additional Configuration Command Responses
{STX} “Z” xx yy zz CCITT/CRC {ETX} breath byte string
0x27 0x0C 0x00 0x1A {STX}Z270C001A{ETX}
02h 5Ah 32h37h 30h43h 30h30h 31h41h 03h
Host Command Module Response Description
<Axx>
Change breath data clear rate
<axx> response = echo data field
<Nxx>
Change waveform update rate
<nxx> response = echo data field
<Oxx>
Change breath data update rate
<oxx> response = echo data field
<Qxxy>
Function enable
<qxxy> response = echo new/current status
<F20yy>
Change pump flow rate
<F20yy> response = echo data field
<C80>
Request software reset
no response