LC101 CO
2
Module OEM Implementation Manual
Page 74 Confidential Welch Allyn OEM Technologies
• Check the exhaust tubing for an improper connection to a vacuum source.
• Refer to Section 5, Pneumatic Operation, for additional information.
84, Barometric Pressure Too High
Barometric pressure transducer reading is greater than the upper barometric limit.
• Check that the host monitor is operating at barometric pressures less than 824 mmHg
(1,300 ft below sea level).
•Possible transducer error.
85, Barometric Pressure Too Low
Barometric pressure transducer reading is less than the lower barometric limit.
• Check that the host monitor is operating at barometric pressures greater than 408
mmHg (15,500 ft above sea level).
•Possible transducer error.
Comentarios a estos manuales