Host Command/Module Response Overview Table
Command Command Description Response Response Description
<M21> Enter standby mode <Sxx06> Acknowledge with Module status
<M23> Enter measurement mode <Sxx06> Acknowledge with Module status
<M24> Enter auto run mode <Sxx06> Acknowledge with Module status
(Invoked with enter measurement
or auto run mode command)
<Wxxyy> CO
2
waveform data packets sent at
programmed interval
(Invoked with enter measurement
or auto run mode command)
<Zxxyyww> Breath data packets sent at pro-
grammed interval
<C00> Request Module status <Sxxyy> Acknowledge with Module status
<C10> Sample Zero Reference Gas <Sxx25> Status 25 = Cal in Progress
<C11> Sample Span Reference Gas <Sxx25> Status 25 = Cal in Progress
<C12> Perform Zero Calibration <Sxx25> Status 25 = Cal in Progress
<C13> Abort Calibration <Sxx00>
<Sxx22>
Status 00, ok (operation aborted)
Calibration not in Progress
<C20> Request software version <Vxxxmmd-
dyyyy>
Version # with month, day, year sent
<C21> Request hardware version <Hxx> Hardware version # sent
<C22> Request ambient barometric
pressure
<Lxxxx> Ambient barometric pressure sent
<C23> Request sensor temperature <Txx> Sensor temperature sent
<C26> Request single CO
2
packet <wxxyy> CO
2
measurement sent
<C27> Request single breath data
packet
<zxxyyww> Breath data packet sent
<C2E> Request current pump flow rate <Bxxxx> Current pump flow rate sent
<C31> Request sensor EEPROM revi-
sion #
<Rxx> Revision # sent
<C32> Request customer code <Ixx> Sensor customer code sent
<C33> Request sensor serial # <Nxxxx> Sensor serial # sent
<C3A> Request last calibration date <Dmmd-
dyyyy>
Last calibration date sent
<C80> Software reset No response
<Axx> Specify breath data timeout <axx> Data field echoed
<F20yy> Specify pump flow rate <f20yy> Command echoed
<Gxx> Update Span Gas <Sxx24> Status 24 = Cal Ready for Next Step
<Lmmd-
dyyyy>
Update Calibration Date. <Sxx24> Status 24 = Cal Ready for Next Step
<Nxx> Specify CO
2
waveform data
packet interval
<nxx> Data field echoed
<Oxx> Specify breath data packet inter-
val
<oxx> Data field echoed
<Qxxy> Specify measurement compen-
sation ON or OFF, SW reset
<qxxy> New and previous measurement
compensation settings echoed