Module OEM Implementation Manual
Page 40 Confidential Welch Allyn OEM Technologies
Calibration Command Responses
Refer to the Status Message Code Table and the Host Command/Module Response Overview
Ta b le at the end of this section for message descriptions.
Host Command Module Response Module Response Format
<Lmmddyyyy>
Update calibration date
<Sxxyy> xx = mode
yy = message
Example:
<S6424>
64= Autorun mode
24= Cal. ready for next step
Example:
<S6421>
64= Autorun mode
21= Cal. already in progress
<Gxx>
Update span gas concen-
tration
<Sxxyy> xx = mode
yy = message
Example:
<S6424>
64= Autorun mode
24= Cal. ready for next step
Example:
<S6422>
64= Autorun mode
22= Cal. not in progress
<C10>
Sample zero reference gas
<Sxxyy> xx = mode
yy = message
Example:
<S6425>
64= Autorun mode
25= Cal. in progress
<C11>
Sample span gas
<Sxxyy> xx = mode
yy = message
Example:
<S6425>
64= Autorun mode
25= Cal. in progress
<C13>
Abort calibration
<Sxxyy> xx = mode
yy = message
Example:
<S6400>
64= Autorun mode
00= OK (operation aborted)