Welch-allyn SureTemp PlusModule - User Manual Manual de usuario Pagina 68

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 85
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 67
Typical Start-up Sequence SureTemp Plus Module OEM Implementation
68 Confidential Welch Allyn OEM Technologies
7. The module sends unsolicited <941C> Temperature Record packets as soon as the probe is
removed from the well.
8. The host monitors the state byte in the Temp packets. When the module goes from Not
Ready to Ready, the host can sound an audible tone or beep to signify that the oral/axillary
probe is ready to be inserted into the patient’s mouth. At this point, the probe is inserted. If
the probe is not inserted, the module remains in predict mode for one minute, and then
moves to the Monitor mode. It remains in Monitor mode until the cycle is inactive.
9. The host monitors temp state and error message bytes within the <941C> Temperature
Record packets.
10. A Final Predict Temperature is included by the module in the temp packet. The <941C> Temp
Record packets continue until the Probe is returned to the well or times out.
Vista de pagina 67
1 2 ... 63 64 65 66 67 68 69 70 71 72 73 ... 84 85

Comentarios a estos manuales

Sin comentarios