SureTemp Plus Module OEM Implementation Communications Protocol
Welch Allyn OEM Technologies Confidential 47
(3604) Set Temperature Probe
Causes the unit to simulate the removal or replacement of the temperature probe.
Command Code (hex)
3604
Data
Size (bytes) Type Description
1 boolean 0: Probe in holder (not ready for reading)
1: Probe removed from holder (ready for reading)
Response Data
Size (bytes) Type Description
0 N/A None
Notes
To use this command, the temperature probe should normally be locked.
(9401) Send Thermometer Status No Clear
Causes the module to return a status block. The status block indicates what the module is
doing and reports any error conditions.
Command Code (hex)
9401
Data
Size (bytes) Type Description
0 N/A None
Response Data
Size (bytes) Type Description
4 TEMP_STATUS Status block
see TEMP_STATUS (page 34) for details
Notes
This command does not clear POST errors.