SureTemp Plus Module OEM Implementation Communications Protocol
Welch Allyn OEM Technologies Confidential 59
(2A12) Set Temperature Method
Causes the unit to read the current temperature method from the host. This specifies whether
the device enters into normal predictive or monitor mode when a probe is pulled.
Example:
Command String 03 02 10 2A 12 01 01 00 01 AC
Preamble 03 02 10
Command Code 2A 12
Type code 01 (command being sent)
Data size 01 00 (1 byte being sent)
Data 01 (1 byte where 01 = Monitor, 00 = Normal predict)
Checksum AC
Response String 03 02 10 2A 12 02 00 00 AD
Preamble 03 02 10
Command Code 94 34
Type Code 02 (good response)
Data Size 00 00 (0 bytes returned)
Data none
Checksum AD
Comentarios a estos manuales