SureTemp
®
Model 678/SureTemp
®
Model 679 v
SPECIFICATIONS
•
Case Dimensions
(nominal):
2.25 in. x 3 in. x 7 in
5.7 cm x 7.6 cm. x 17.8 cm
•
Case Material:
ABS Plastic
•
Weight (nominal):
10.25 ounces with batteries and probe
•
Input:
Welch Allyn Thermistor Probe. P/N 02678-000
(ORAL 4’), P/N 02678-100 (ORAL 9’), 02679-000
(RECTAL 4’), AND 02679 -100 (RECTAL 9’)
•
Display range:
28.9
o
C to 42.2
o
C (84.0
o
F to 108.0
o
F)
•
Laboratory Accuracy:
± 0.2
o
F in the Monitor mode and in a water bath
per Welch Allyn document number 90565-000
•
Clinical Accuracy:
Meets the ASTM E1112-86 clinical test standard
•
Push buttons:
Mode and Pulse Timer
•
Mode Button:
Selects
o
F/
o
C, Oral, Axillary and Monitor Modes
•
Probes:
Interchangeable Oral (also used for Axillary) and
Rectal
•
Power Source:
Three “AA” Batteries
•
Battery Operating Life:
Approximately 5,000 temperature measurements
(At 72
o
F ambient temperature)
•
Display Type:
Liquid Crystal Display, 3½ digits plus special icons
•
Operating temperature:
16
o
C to 40
o
C (60.8
o
F-104
o
F) @ 15% to 95%
RH non-condensing per ASTM E1112-86
•
Storage Temperature:
-20
o
C to 50
o
C (-4
o
F-120
o
F) @ 15% to 95%
RH non-condensing per ASTM E1112-86
Comentarios a estos manuales