3. CALIBRATION TESTING
The Calibration Key provides a convenient means of testing the thermometer.
Calibration Key Procedure
1. Extract the probe and disconnect it from the thermometer.
2. Insert the Calibration Key (part number 01637-000) into the probe connector
receptacle on the thermometer and observe the display. The display should read
CAL for two seconds, and then go blank.
3. Insert a probe shaft into the probe storage well and remove it to initiate a
temperature taking cycle. Wait for the display test and then observe the display.
The display must read between 97.1
°
F and 97.5
°
F inclusive for the calibration of
the instrument to be correct.
4. Remove the Calibration Key and reinstall the probe connector plug.
5. Then install the probe into the probe storage well.
Note:
This Cal Key test does
not
test the probe. To do so requires the use of the
Welch Allyn Model 9600 Calibration Tester.
If the reading from the Cal Key is not within the specified range or you are having
other problems with the use of the Cal Key, refer to
Cal Key Problems
in the
Troubleshooting
section.
Model 9600 Calibration Testing Procedure
The Model 9600 Calibration Tester provides a convenient means of testing the entire
thermometer system (instrument and probe).
The 9600 must be warmed up and stable at one of the two available temperature
settings.
The thermistor based instrument under test must be in Monitor mode and no probe
cover loaded.
The probe is inserted into the small hole in the dry heat well of the 9600 and allowed
to settle for a minimum of 2 minutes to the final temperature. The reading on the
thermometer must be within the range specified on the 9600. Refer to the
Model 9600
Operation Manual
for complete instructions.
Note:
All Welch Allyn and Diatek thermometers (thermistor and infrared ear
thermometers) can be checked in the Model 9600.
If you are having problems with the use of the Model 9600, refer to the
Troubleshooting
section in the
Model 9600 Operation Manual
.