4
33
Troubleshooting
About the connectivity software modules
The Connex CSK software comprises a set of modules that are installed separately as
needed. The connectivity modules are listed here:
Firewall settings
If you are using a firewall between any PCs in your configuration, the following ports or
application files are required to gain access through the firewall. Software installation
automatically opens these ports.
Module Full name Description Installation location
CPSAPI CardioPerfect
Server
Application
Programming
Interface
This is the database connectivity module. It plays a
central role, directing communications between
various software modules and the server software.
One CPSAPI per site
DC Data Catcher This module receives data from the device(s). One DC per CPSAPI installation
and per Workstation that supports
wired-USB
DCP Device
Communication
Protocol
This module listens for device(s) that are
connected wirelessly, and it tells them which IP
address to communicate with.
One DCP per subnet
FI File Importer This module receives tests from removable media
(SD memory cards or USB storage devices).
One FI per Workstation receiving
tests via removable media
Software module Port Type Application file
CPSAPI 9247 TCP CPWS.Service.exe
DCP 44435 UDP DCPd.exe
DC 281 TCP DataCatcher.exe
CP 200 ICMP Allow incoming echo requests Ping test
CP 20 4242 TPC/IP WACP20.exe