26 Configuration Welch Allyn Connex CSK software
General settings
1. Stop button Selecting this stops the device server portion of Data Catcher. All state information pertaining
to currently pending information host messages will also be reset.
2. Start button Selecting this starts the device server portion of Data Catcher.
3. Automatic
configuration
This option allows you to fully configure Data Catcher using an existing configuration file. This
will copy the specified file to the Data Catcher directory, rename the file to
“DataCatcherConfig.xml,”and force the application to restart to reload the new configuration.
Note: This will disconnect any current devices in the same way the Stop button works. This will
also restart the Web interface portion of the application to employ any user interface changes.
4. Local data
cache directory
Data Catcher provides the option to save all device data sent through the system to a local
directory. This data is the unaltered data, directly from the device. This configuration option
specifies the directory where the cached device data files should be stored, relative to the
application directory.
5. Local data
cache file types
This option allows you to specify what types of data files should be cached locally by Data
Catcher. You can choose to save the pure binary data from the device, an XML form of the data,
or both. Each file is uniquely named using a combination of the device source ID, the current
system time, and a randomly generated number.
6. Save settings Selecting this commits any changes made to the configuration settings. This will update the
application settings on the fly, and save the changes to the “DataCatcherConfig.xml” file.