TEST
Figure 10
The TEST message indicates that the test has started. Immediately following
this message, test data will begin to appear.
BLOCK
Figure 11
The test cannot continue since the measured admittance is less than 0.2 mmho.
Possible causes:
probe tip is lodged against canal wall
ear canal occluded
collapsed ear canal
extremely small ear or tortuous canal
probe tip is clogged with cerumen
Possible solutions:
reposition the probe tip
perform otoscopy to check for occlusion
remove cerumen from probe tip
10
-400 -200 0 200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc
-400 -200 0 200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc