Mainstream CO
2
Sensor
Mainstream NDIR single-beam, single
path/wavelength, ratiometric
Warm-up time (CO
2
sensor and monitor):
45 sec typical, 3 minutes maximum
Automatic zeroing
No routine calibration
Operating altitude: -2,000 to 15,000 ft
(817 to 429 mmHg)
Dimensions: 1.003 in (H) x 1.036 in (W) x 0.78 in (L)
(2.5 x 2.6 x 2.0 cm)
Weight: 12 g (< 0.43 oz)
Cable length: 10 ft (3.05 m) nominal
CO
2
Airway Adapters
Disposable or multi-use
Size: 15 mm ID (meet ISO specifications)
Single or multi-use adult/pediatric airway adapter
for patients who weigh < 5 kg:
Added deadspace: < 6 cc (0.37 cubic in)
Single use low deadspace airway adapter
for
patients who weigh < 5 kg:
Added deadspace: < 0.6 cc (0.037 cubic in)
Wireless Option
FlexNet
Network: 2.4 GHz Wireless Local Area
Network (WLAN) and 10/100 Base-T Ethernet
network
Frequency: 2.402 to 2.480 GHz (USA)
Modulation: GFSK, Frequency Hopping Spread
Spectrum (FHSS)
Output power: 112 mW (maximum)
IEEE 802.11 compliant
Wireless Propaq CS monitors per Access Point:
7 (max.)
Nurse Call Option
Maximum switch current: 1A
Maximum switch voltage: 30 V AC/DC
Isolation: 1500 Vrms
Alarm relay: Energized during apnea alarm or
patient alarm
Customized cable: One end is a 4-pin plug,
compatible with the monitor’s nurse call
connector; the other end must be customized to
connect to the local nurse call system
Printer Option
Thermal sensitive dot
Numeric annotation: date, time, all active
parameters, patient name if connected to Acuity
Number of waveforms printed simultaneously:
up to 3
Print speeds: 6.25, 12.5, 25.0 mm/sec
Format: 53 mm wide print area on 60 mm wide paper
Operating modes:
Continuous: start/stop real-time printouts with
all active numerics and up to 3 waveforms
Snapshot: 8 sec (32 sec for CO
2
/Resp) of
immediate history for all active numerics and up
to 3 waveforms
Alarm print: 20 seconds total/12 seconds prior
to alarming parameter
Auto print: Snapshot printout every 15 or 30
min, or 1, 2 or 4 hours
Trend print: On demand or Auto Trend every
4 hours. Up to 8 hours in tabular format.
OxyCRG trend print: Print-only function (CO
2
or
Resp must be installed): 2 min printout of trended
HR/PR and SpO
2
numerics and compressed CO
2
or Resp waveforms (CO
2
is priority source), with
annotation and MIN/MAX tabular report
OxyCRG on alarm: Set to print out 60 sec after
HR/PR or SpO
2
alarm or 75 sec after BR/RR or
apnea alarm
Apnea ticket: Auto print HR/PR, SpO
2
and
elapsed time after apnea event and after 1
minute of continued apnea
Cuff ticket: Auto print NIBP, HR/PR, SpO
2
, CO
2
and RR/BR after each NIBP reading
Alarms
All parameters: upper/lower limits
All parameters: Adult/Pediatric/Neonate patient
mode-specific limits
Factory default or programmable settings for
all patient modes
Alarm indicator: red
Alarm(s) off indicator: yellow
Audible alarm tone: high/med/low
Alarm suspend: 90 seconds
One-button Stat Set for all alarm limits
(except Apnea Delay)
EQ Alert: yellow
Trends
Tabular numeric format
All parameters trended/viewable
Resolution: non-NIBP trends entered every
2 min
NIBP trends entered after each reading
Duration: 5 hours non-NIBP trends
(up to 150 readings)
8 hours NIBP trends (up to 128 readings)
Page up/down trend view
Inservice Mode
Includes two sets of simulated patient data
including waveforms, for training and education
Color Active Matrix Display
Type: TFT (Thin Film Transistor) LCD module
Resolution: 640 x 480 pixels, 1 pixel = R+G+B
dots
Active viewing area: 6.73 x 5.10 in
(170.9 x 129.6 mm)
Pixel pitch: 0.0105 in (0.267 mm)
Viewing angle: U/D 40°, R/L 60° (typical),
10:1 contrast ratio
Contrast ratio: 150:1 (typical); measured in dark
room at center of screen
Display color: 18-bit (6 bits per primary color)
Waveforms and matching numerics: Predefined
Luminance: 200 cd/m
2
(typical); measured at
saturation point
Response time: 40 ms (maximum); “white
to black”
Environmental
Operating temperature: 5 to 40
°
C (monitor),
5 to 40 °C (printer)
Operating relative humidity: 15% to 95%,
noncondensing (MIL STD 810E), printer 35%
to 85%, noncondensing (MIL STD 810E)
Operating altitude: -2,000 to 15,000 ft
(-610 to 4,572 m)
Shipping/Storage temperature: -20 to 60
°C
Shock: 50 g (monitor); 30 g (expansion
module/printer)
Random vibration: Designed to meet RTCA
DO-160D, category C (monitor)
Electromagnetic compatibility (EMC): per
EN60601-1-2, 1993
Drip-proof per IEC 529, level IPX1 (monitor)
Physical
Monitor (including handle):
Size: 8.2 in (H) x 9.6 in (W) x 5.6 in (D)
(20.8 x 24.4 x 14.1 cm)
Weight: 7.6 lb (3.4 kg)
Monitor with SpO
2
Module:
Size: 8.2 in (H) x 9.6 in (W) x 7.7 in (D)
(20.8 x 24.4 x 19.7 cm)
Weight: 10.8 lb (4.9 kg)
Monitor with Expansion Module
(with Printer/SpO
2
/MSCO
2
)
Size: 11.4 in (H) x 9.6 in (W) x 7.7 in (D)
(28.8 x 24.4 x 19.7 cm);
Weight: 14.4 lb (6.5 kg)
Power
Rechargeable sealed lead acid internal
battery pack
Internal recharger
Input voltage: 12 to 28 VDC, 25 W
AC power adapter: 100 to 120 VAC, 50 to 60 Hz
(North America/Japan); 220 to 240 VAC, 50 to
60 Hz (International)
Operating time on battery, typical*: Monitor
with SpO
2
option: 4 hours; Monitor with
Expansion Module (Printer, SpO
2
/CO
2
): 3 hours;
Monitor only: 2 hours
Battery recharge time: 6 to 8 hours
(when monitor OFF)
Battery recharge time: 8 to 12 hours
(when monitor ON)
Certifications/Standards
Complies with relevant AAMI, IEC, EN, CSA
and UL standards. CE Marked according to
the European Medical Device Directive. CSA
Certificate of Compliance for use in the U.S.
and Canada
VA Contract V797P-3486k
DSCP Contract SP0200-97-D-8021
* Fully charged, new battery at 25 °C, NIBP and
Snapshot (if applicable) every 15 minutes, all
accessories used.
Smartcuf, Propaq and Acuity are registered trademarks of
Welch Allyn. Other product and company names mentioned
herein are for identification purposes only and may be the
trademarks or registered trademarks of their respective
companies. Specifications subject to change without notice.
© 2006 Welch Allyn SM2517 Rev C
The Propaq CS is configurable with:
Heart/Pulse rate*
3- or 5-lead ECG*
Motion-tolerant noninvasive blood pressure*
Temperature (2 channels)*
Invasive blood pressure (up to 2 channels)
Masimo
®
SET Motion-Tolerant pulse oximetry*
Nellcor OxiMax
®
pulse oximetry*
Capnography (mainstream and/or sidestream)
Impedance respiration/apnea*
Acuity
®
Central Station interface (wireless/hardwire)
Nurse call interface
*Standard Features
The Welch Allyn
®
Propaq
®
CS
Vital signs monitoring where you
need it, when you need it.
Propaq
®
CS Specifications
All Propaq vital signs monitors come standard
with Smartcuf technology, a new standard in NIBP
accuracy. This patented software offers industry-
leading performance and accuracy in the presence
of motion artifact.
®
Only Masimo SET SpO
2
sensors should be
used with Masimo pulse oximetry. Masimo
and Masimo SET are registered trademarks
of Masimo Corporation.
Only Nellcor
®
OxiMax
®
pulse oximetry sensors
should be used with the Nellcor pulse oximetry
option. Nellcor and OxiMax are registered
trademarks of Nellcor Puritan Bennett, Inc.
8500 S.W. Creekside Place
Beaverton, OR 97008-7107
phone 503.530.7500
fax 503.526.4200
toll-free 80
0
.289.2500
technical support 800.289.2501
www.welchallyn.com
MC 3146 SM2517RevC Propaq CS 2/1/06 9:37 AM Page 2
Vista de pagina 6
1 2 3 4 5 6 7 8

Comentarios a estos manuales

Sin comentarios