ECG_Sync Requirements
ECG_SYNC connection requirements are shown in Figure 6. Refer to the POEM OEM
Implementation Manual for timing requirements and instructions about selecting the polarity of
ECG_SYNC.
Figure 6 - ECG_SYNC Connection
Ground of host system.
ECG_SYNC from host:
(levels = 0 V and +3.3 V to +5 V)