Welch-allyn Network Installation Best Practices Overview - Ins Manual de usuario Pagina 8

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 10
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 7
4 Best practices Welch Allyn Network installation
Critical for Connex CS and Connex RMS
Best practice Affected types Without best practice
Ports allowed
TCP: 281, 283
UDP: 291, 7711-7719, 44435-44436 (only if Spot LXi is used
with Connex)
Wired Connections cannot be established.
Rendezvous Perform at least one of the following:
Allow UDP broadcasts on ports 7711-7719, or 44435-44436
(only if Spot LXi is used with Connex)
DNS name resolution for Connex servers using a locally
configured name
Configure the device with a fixed IP address of the Connex
server
Wired Connections cannot be established from the Welch
Allyn VLAN to the server.
SSID/Radio settings
a band Only
Wireless Loss of connection and data, patient monitor will not
connect.
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10

Comentarios a estos manuales

Sin comentarios