Welch-allyn Mach 6 Quick Reference Guide - Quick Reference Gui Manual de usuario

Busca en linea o descarga Manual de usuario para Equipo Welch-allyn Mach 6 Quick Reference Guide - Quick Reference Gui. Welch Allyn Mach 6 Quick Reference Guide - Quick Reference Guide User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 2
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
MACH 6 Quick Reference Guide
K990291 Rev. C
Daily Requirements Powder Canister Level and moisture
Monthly Requirements Filters and Moisture
Handpiece/s Handpiece Hose/s and Nozzles
Pinch Tube (check)
Quarterly Requirements Pinch Tube (replace)
BI-Annual Requirements Pinch Valve Assembly (lubricate)
Inspect Canister components
Annual Requirements Air In and Out In-Line Filters (replace)
Internal Tubing (re-tube)
Powder Canister (replace components)
Maintenance Checks
No. Maintenance item 6 mo. 12 mo. 18 mo. 24 mo.
PM 1 Maintain Pinch Valve XX XX XX XX
PM 2 Replace Filters & Tubing XX XX
PM 3 Maintain Powder Canister XX XX
PM 4 Calibrate system XX XX XX XX
PM 1. Maintain pinch valve block 6/12 Months): .50 hour
One Year Maintenance Kit K150034
PM 2. Replace the Filters and tubing (12 months): 1.0 hour
One Year Maintenance Kit K150034
PM 3. Check Powder Canister (6/12 months): 0.5 hour
One Year Maintenance Kit K150034
Check Canister Gasket and lid
Clear powder ports and back flush canister
Clean Gasket and Canister lid
Clear and measure Orifice Plate holes
Replace powder and verify system operation
Maintain Powder Canister (12 month) 1.0 hour
One Year Maintenance Kit K150034
PM 4. Calibrate system (6 months): 1.0 hour
Verify and calibrate using procedure
Vista de pagina 0
1 2

Indice de contenidos

Pagina 1 - Maintenance Checks

MACH 6 Quick Reference Guide K990291 Rev. C Daily Requirements Powder Canister Level and moisture Monthly Requirements Filters and Mo

Pagina 2 - Troubleshooting Guide

MACH 6 Quick Reference Guide K990291 Rev. C Troubleshooting Guide PROBLEM SOLUTION No powder or air from the handpiece nozzle. ?? The handpiece

Comentarios a estos manuales