Welch-allyn CP 200 Spirometry Option - User Manual Manual de usuario Pagina 17

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 92
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 16
Directions for Use Chapter 2 Reviewing the Spirometry Settings 13
Reviewing the Operation Settings
To review or change the settings that affect the overall operation of the spirometer,
reflected both on screen and in print, follow these steps.
1. Press the Menu key .
2. Choose Spirometry Settings > Operation Settings.
The following screen appears.
Figure 2. “Spirometry Operation Settings” Screen
3. If desired, change the settings.
Setting Description
Select Protocol The selected protocol determines the way the spirometer operates when testing a patient.
Applicable for FVC testing only. For details, see “Spirometry Protocols” on page 55.
•None
PCP (primary care practitioner)
•NIOSH
OSHA
SSD (Social Security & Disability)
Select Adult
Predictive Norm
The selected adult norm is the primary source of predictive values for adult patients.
For details, see “Norm Profiles” on page 66.
Select Ped.
Predictive Norm
The selected pediatric norm is the primary source of predictive values for pediatric patients.
For details, see “Norm Profiles” on page 66.
Select Best Effort
Formula
A patient’s best effort is a measurement calculated from a set of efforts. To determine the
way in which best effort is calculated, choose from these options:
Best Measurement —
Defines best effort as the single best effort in a set of efforts (best FVC-pre, best FVC-
post, best SVC). This ATS-recommended method uses the effort with the highest sum
of FVC + FEV1, or the effort with the highest SVC value. (For details, see the document
noted in Reference 6 on page 75.)
Best Composite —
Defines best effort as a composite of the highest parameter values across all selected
efforts (except FVC and FEV1, which are both selected from the highest sum of
FVC + FEV1.)
Spirometry Operation Settings
9:17AM Oct 16 08
1 Select Protocol
2 Select Adult Predictive Norm
3 Select Ped. Predictive Norm
4 Select Best Effort Formula
5 Select FVC Reversibility Formula
6 Select FEV1% Formula
7 Enable Predictive Points
8 Enable Predictive Curve
9 Enable ATS Interp. Results
A Enable Composite Norm Values
Vista de pagina 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 91 92

Comentarios a estos manuales

Sin comentarios